E-20C-H-O / E-20CJ-H-O User Manual V2.1 Table of Contents 1. Symbols ........................1 1.1 Control Buttons ················································································ 1 1.2 Device Symbols ················································································ 1 2. Warning, Caution and Important Tip ................2 3. Intended Use ......................2 4. Contraindications ..................... 3 5.
Page 3
E-20C-H-O / E-20CJ-H-O User Manual V2.1 17. Transferring the Device to Another Patient ............. 25 18. Reordering ......................25 19. Technical Support ....................26 20. Disposal ......................26 21. Troubleshooting ....................26 21.1 Common Problems in Patients and Corresponding Solutions ······················· 27 21.2 Common Problems in the Device and Corresponding Solutions ···················...
E-20C-H-O / E-20CJ-H-O User Manual V2.1 1. Symbols 1.1 Control Buttons Ramp Button Mute Button Knob 1.2 Device Symbols Follow Instructions for Use Operating Instructions Type BF Applied Part (mask) Class II (Double Insulated) AC Power DC Power IP22 ≥12.5 mm Diameter, Dripping (15ºtilted)
E-20C-H-O / E-20CJ-H-O User Manual V2.1 2. Warning, Caution and Important Tip WARNING! Indicate the possibility of injury to the user or operator. CAUTION! Indicate the possibility of damage to the device. IMPORTANT TIP! Place emphasis on an operating characteristic.
E-20C-H-O / E-20CJ-H-O User Manual V2.1 4. Contraindications Studies have shown that the following pre-existing conditions may contraindicate the use of positive airway pressure therapy for some patients: Absolute Contraindications: pneumothorax, mediastinal emphysema; cerebrospinal fluid leak, traumatic brain injury, or pneumocephalus; shock caused by a variety of conditions before treatment;...
E-20C-H-O / E-20CJ-H-O User Manual V2.1 5. Specifications Device Size Dimensions: 170 mm × 180 mm × 118 mm, or 290 mm × 180 mm × 134 mm (with the humidifier) Weight: 1.5 kg, or 2.5 kg (with the humidifier)
E-20C-H-O / E-20CJ-H-O User Manual V2.1 Ramp The ramp time ranges from 0 to 60 minutes. Sound Pressure Level <30 dB, when the device is working at the pressure of 10 hPa. Sound Power Level <38 dB, when the device is working at the pressure of 10 hPa.
This feature is only available for E-20C-H-O. Auto On With this feature, the device automatically initiates therapy when you breathe into the mask. This feature is always enabled. This feature is only available for E-20C-H-O. CPAP Continuous Positive Airway Pressure.
• This device should only be used with the mask and accessories manufactured or recommended by BMC or with those recommended by your prescribing physician. The use of inappropriate masks and accessories may affect the performance of the device and impair the effectiveness of therapy.
• When the insulation layer of the SpO probe cable is damaged, do not connect the probe to the patient. • Please contact BMC to buy the SD card if you need it. 10. System Features Fig. 10-1 Name Function Indicate the humidity level.
E-20C-H-O / E-20CJ-H-O User Manual V2.1 Fig. 10-2 Name Function SD Card Slot Insert the SD card into this slot Communications Connected to external equipment (Not for connection to the phone Port or computer) DC Inlet An inlet for the DC power supply...
E-20C-H-O / E-20CJ-H-O User Manual V2.1 11.2 Installing the Air Filter and Filter Cap (1) Attach the air filter to the filter cap, as shown in Fig. 11-1. Fig. 11-1 (2) Install the filter cap containing the air filter to the main device, as shown in Fig. 11-2.
E-20C-H-O / E-20CJ-H-O User Manual V2.1 CAUTION! • Inspect the power cord often for any signs of damage. Replace a damaged cord immediately. IMPORTANTS! • After interruption and restoration of the power supply, the device will restore its pre-interruption working status automatically.
E-20C-H-O / E-20CJ-H-O User Manual V2.1 - Use the accompanying tube and mask provided by BMC. - Do not wear the mask for more than a few minutes while the device is not operating. - Use only masks with vent holes. Do not block or try to seal the vent holes in the exhalation port.
E-20C-H-O / E-20CJ-H-O User Manual V2.1 Fig. 11-7 If the SD card is inserted incorrectly or not inserted, a symbol indicating incorrect insertion or no SD card present will appear in the Main Interface on the screen of the device, as shown in Fig.
E-20C-H-O / E-20CJ-H-O User Manual V2.1 Fig. 11-10 (2) Point the four holes of the Base Plate towards the four reference points on the back of the main device to properly stick the plate to the device, as shown in Fig. 11-11.
Page 18
E-20C-H-O / E-20CJ-H-O User Manual V2.1 Communications Port. The model of the SpO Kit is KS-CM01. The SpO Kit is calibrated to display FUNCTIONAL OXYGEN SATURATION. Attach its sensor to the patient's index finger or any other finger. The sampling rate of the SpO signal is about 50 Hz, and the update rate of the frame is 1 Hz.
• Be sure to follow your physician’s instructions on adjusting the settings! To order any accessories not included with this device, contact your equipment supplier. • DO NOT connect any ancillary equipment to this device unless recommended by BMC or your physician. If you suffer from chest discomfort, shortness of breath, stomach bloating, or severe headache when using the device, contract your physician or qualified medical personnel immediately.
E-20C-H-O / E-20CJ-H-O User Manual V2.1 CAUTION! • Observe the water level of the water chamber before using the humidifier. Make sure there is sufficient water in the water chamber, and avoid heating the humidifier with an empty water chamber.
E-20C-H-O / E-20CJ-H-O User Manual V2.1 13. Navigating the Patient Menu 13.1 Steps to Navigating the Patient Menu 13.1.1 Accessing the Main Interface Connect the power cord and power adapter properly. The screen displays the Main Interface shown in Fig. 13-1.
E-20C-H-O / E-20CJ-H-O User Manual V2.1 13.1.4 Selecting Options As you turn the Knob clockwise, the cursor moves downwards from one option to another. As you turn it counterclockwise, the cursor moves upwards. When the cursor is on a certain option, press the Knob , and the option is then displayed in yellow, meaning that the option can now be adjusted, as shown by the Humidifier option in Fig.
E-20C-H-O / E-20CJ-H-O User Manual V2.1 Fig. 13-7 Note: are page turning symbols. 13.1.8 Exiting the Patient Menu (1) Returning to the Initial Setup Interface Move the cursor to the Back option by turning the Knob , as shown in Fig. 13-8.
Type Pillow; choose other suitable masks as well. When selecting masks Other other than the above three types of BMC masks, the patient can identify the masks as other iCode provides access to the patient's compliance data iCode, during a recent time period. The iCode QR mode displays...
E-20C-H-O / E-20CJ-H-O User Manual V2.1 14. Alert Alert Message Description An audible alert will sound if the device is accidentally disconnected from power when it is delivering air. Note: Power Failure!!! (1) The alert will not sound if power failure occurs when the device is in standby state.
E-20C-H-O / E-20CJ-H-O User Manual V2.1 15. Cleaning and Disinfection WARNINGS! • Regular cleaning of the device and its accessories is very important for the prevention of respiratory infections. • To avoid electric shock, always unplug the device before cleaning.
E-20C-H-O / E-20CJ-H-O User Manual V2.1 water thoroughly. (3) After cleaning, air-dry the tube in a cool, well-ventilated area, and avoid direct sunlight. It takes approximately 30 minutes to completely air-dry the tube. Check whether the tube is completely dry before reuse.
• If the device is used when the atmospheric pressure is out of the stated range (See Section 5), the accuracy of the leakage alert will be affected. (1) Use the BMC carrying case to carry the device and accessories along with you. Do not put them in your checked baggage.
19. Technical Support Please contact BMC directly if you need the circuit diagram of the device and the list of components for certain purposes such as maintenance or connection to other equipment. BMC will provide the circuit diagram and / or other technical documents in whole or in part according to your needs.
E-20C-H-O / E-20CJ-H-O User Manual V2.1 21.1 Common Problems in Patients and Corresponding Solutions Problem Possible Cause Solution(s) Increase the humidity setting The nose reacts to the airflow of the humidifier. Dry, cold, runny, and cold. Due to fast airflow, the Contact your physician, and and blocked nose;...
Page 31
E-20C-H-O / E-20CJ-H-O User Manual V2.1 Turn humidity setting down, or raise the room temperature. When the humidifier is used, the Place the tube under the quilt, humidified air tends to condense or use the tube cover. Water in mask...
E-20C-H-O / E-20CJ-H-O User Manual V2.1 21.2 Common Problems in the Device and Corresponding Solutions Problem Possible Cause Solution(s) Take a few deep breaths with The Auto On / Off feature is the mask on, and the device will enabled...
E-20C-H-O / E-20CJ-H-O User Manual V2.1 22. EMC Requirements Guidance and manufacturer's declaration - electromagnetic emissions The device is intended for use in the electromagnetic environment specified below. The user of the device should ensure that it is used in such an environment.
Page 34
E-20C-H-O / E-20CJ-H-O User Manual V2.1 Guidance and manufacturer's declaration - electromagnetic immunity The device is intended for use in the electromagnetic environment specified below. The user of the device should make sure that it is used in such an environment.
Page 35
E-20C-H-O / E-20CJ-H-O User Manual V2.1 Guidance and manufacturer's declaration - electromagnetic immunity The device is intended for use in the electromagnetic environment specified below. The user of the device should make sure that it is used in such an environment.
Page 36
E-20C-H-O / E-20CJ-H-O User Manual V2.1 Recommended separation distances between portable and mobile RF communications equipment and the device The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent...
Page 37
E-20C-H-O / E-20CJ-H-O User Manual V2.1 Recommended separation distances between wireless communications equipment The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent...
(1) year for main device and three (3) months for all accessories from the date of sale by BMC Medical Co., Ltd. to the dealer. If the product fails to perform in accordance with the product specifications, BMC Medical Co., Ltd. will repair or replace, at its option, the defective material or part.