Abbott FreeGo Training Manual page 22

Pump
Table of Contents

Advertisement

TROUBLESHOOTING GUIDE
SCREEN DISPLAY
FEED
EMPTY
LOCK
PUMP LOCKED
NO
CASSETTE
OCCL
INTO PUMP
or
OCCL
OUT OF PUMP
PUMP
READY TO USE
HOLD MINUTES
If you have any problems setting up your FreeGo® pump and feed system,
please call Abbott Nutrition on Freephone 1800 22 11 66
POSSIBLE CAUSE
This means that air has been
detected in the giving set
Lock-out disables the SET
RATE, SET DOSE and CLEAR
VOL dial settings, as well as the
pump's prime features
This means that either there is
no cassette installed or that the
cassette was installed incorrectly
when attempting to run pump or
prime giving set
Flow has stopped due to
blockage between the feed
container and the pump, or
between the pump and the
patient
The battery charge has reached
the minimum level to operate
and the pump can now be used
This display counts down the
remaining time the pump will
be on hold
ACTION TO TAKE
If you are just beginning a feeding:
• Open the door. Gently press the tubing to
ensure it is fully seated in slots below the
cassette. Close the door firmly.
Or... If there is more food to deliver:
• Turn dial to HOLD to stop the audio alarm
• Remove giving set adapter from your enteral
feeding tube
• Refer to INSTRUCTIONS FOR USE to prime
the air out of the giving set
• Securely re-attach the giving set to your
enteral feeding tube
• Turn dial to RUN to begin feeding
Or... If you are done feeding:
• Turn dial to OFF/CHRG. This message is
displayed because pump was programmed to
feed until container is empty (SET DOSE at
INF DOSE).
Turn dial to SET RATE or SET DOSE.
Deactivate the lock by pressing the Up and
Down arrows simultaneously.
Repeat these steps to reactivate the lock feature
Turn dial to HOLD and insert cassette correctly
and close set door. Turn dial to RUN to start
pump, or attempt to prime the giving set again
Turn control dial to HOLD. Remove giving set from
pump. Clear obstruction; check for kinked tubing or
empty container. Reinstall giving set. If this alarm
appears during priming, attempt to prime the set
again, then connect the feeding adaptor to the
enteral feeding tube. Turn dial to RUN to begin
feeding
Turn dial to any position to clear message.
Confirm all settings are accurate. Turn dial to
RUN to continue feeding
Turn dial to HOLD. If desired, adjust the HOLD
timer to any duration (in 1 minute intervals),
between 1 to 90 minutes by using the Up and
Down arrows
19