ALARMS/ALERTS: OCCL
SILENCING THE ALARM

OCCL

If the display shows OCCL
INTO PUMP and an audio
alarm sounds, the feed
cannot be delivered due to a
blockage between the feed
container and the pump
REMOVE THE GIVING SET FROM THE PUMP AND CHECK
THE FOLLOWING
Is the giving set tubing bent or trapped anywhere?
Is the two-way tap in the feeding position?
Is the clamp open on your feeding tube (e.g. nasogastric/gastrostomy tube)?
Is your feeding tube blocked? To check, set the pump up ready to feed, disconnect the
giving set from the feeding tube and RUN the pump into a suitable container. If feed
comes out and there is no alarm,it means the pump and giving set are working normally,
indicating that your feeding tube may be blocked (contact your healthcare professional).
Turn dial to
HOLD to silence
the alarm
If the display shows OCCL
OUT OF PUMP and an
audio alarm sounds there
is a blockage between the
pump and the patient
15