Testing the Alarm Relays

www.DataSheet4U.com
4-6
Enter the alarm setup menu, as described on page 4-2. To verify that
the alarm relay is functioning properly, the alarm may be operated
manually as follows:
Alarms..Alarm-Feat
Trip Value
Deadband Value
Test
POWER
ESC
Note: The [Trip Type] and [Fail-Safe] options are also available in
the above list box.
The following options are available in the [TEST] submenu:
Normal: sets the alarm relay to its proper state, as dictated by the
current live measurment value.
IMPORTANT:
Always return the alarm relay to this state at the
conclusion of the testing.
Trip: manually triggers the alarm relay to its "ON" state.
Reset: manually returns the alarm relay to its "OFF" state.
Alarms..Alarm-Test
Normal
Trip
Reset
POWER
ESC
The meter now automatically returns to the "Alarm-Features"
prompt. Either press [ESC] until you exit the User Program or proceed
to the appropriate section to continue setting up the alarms.
Use the [ ] and [ ] keys to
select [Test], and press
[ENTER].
ENTER
Use the [ ] and [ ] keys to
select the desired TEST option.
Press [ENTER] when done.
ENTER