Advertisement

Supply voltage
– AC residual
Power consumption
Electrical connection
Connection thread
Material
Weight
Size:
Ambient conditions
For operation
During storage
Ingress protection
CE mark
9 V DC to 30 V DC
®
For HART
communication:
13 V DC to 30 V DC
®
For HART
communication in multidrop installa-
tion: 9 V DC to 30 V DC
<0.5 V
with analogue 4 to 20 mA signal transmis-
ss
sion
<0.2 V
and <2.2 mV
ss
eff
®
tal HART
communication
5.6 W (typical), <7 W (maximum)
Cable diameter 7 to 12 mm,
Cross-section max. 2.5 mm
3/4" NPT or M 25 x 1.5
Stainless steel SS 316L
Approx. 2.2 kg / 4.85 lbs. (without accessories)
Approx. 160 mm x Ø89 mm / 6.3 inch x Ø3.5 inch
o
–40 to +77
C / –40 to +170
700 to 1300 mbar,
0 to 100% relative humidity
o
–40 to +85
C / –40 to +180
700 to 1300 mbar,
0 to 100% relative humidity, non-condensing
IP 65, IP 66 and IP 67, NEMA 4X
Equipment and protective systems intended for use
in potentially explosive atmospheres
(Directive 94/9/EC).