Dräger PIR 7000 Technical Handbook page 38

Infrared gas transmitter
Table of Contents

Advertisement

Accessories
Mounting:
Remove status display PIR 7000/7200,
flow cell PIR 7000/ 7200 or remote test
adapter PIR 7000/7200 from gas
transmitter.
Remove joint seal ring PIR 7000 if
applicable.
1 Carefully insert process cuvette between
the two ridges of the gas transmitter.
Make sure that the sealing ring is fitted
between process cuvette and gas
transmitter. Ensure that the seals and
sealing surfaces are tightly fitted and
clean.
2 Fit holder to the gas transmitter and
fasten using two screws.
3 Screw on connecting bushes.
Connect gas lines to the connecting bushes. During this process, ensure in particular that
the material is compatible with the substance to be monitored.
The gas inlet and gas outlet connection can be freely selected.
Check system for sufficient gas-tightness.
— The response time is dependent on the set volume flow (see "Performance Characteristics
(typical values)" on page 73).
After each installation or removal of the process cuvette, it is absolutely necessary to
calibrate the zero point and sensitivity!
Operating notes
The display of the Dräger PIR 7000/Dräger PIR 7200 depends on the partial pressure of the
measured gas component. Alternating gas pressures or gas pressures which differ from the
atmospheric pressure may be created in the process adapter due to system-related
parameters such as flow rates, nominal diameters of inlets etc.
The maximum allowed internal pressure is 3000 mbar,
the minimum allowed internal pressure is 700 mbar.
38
1
NOTICE
2
3
3