Autoclave - cleaning and maintenance

For CLEANING the appliance wipe inside and outside with a damp cloth. Never immerse the lid and
chamber of the unit in water.
Before start of cleaning or disinfection plug off mains cable and make sure that appliance has cooled
down.
For DISINFECTION of metal surface disinfection liquids with following substances are suitable:
Ethanol, Propyl alcohol, Isopropyl alcohol, Glutaraldehyde, quaternary ammonium compounds.
Please follow the application notes of the manufacturer for the cleaning and disinfection detergents.
Discoloration and stains occur when not using distilled water. These stains may block the pressure
control valve. Therefore, we recommend using distilled water
To remove discoloration, add some citric acid to the aqua dest. and boil for about 10 minutes, do not
close the CertoClav. Empty the chamber immediately.
The steam release valve nozzle should be cleaned regularly with a needle.
Plastic surfaces shall be cleaned with a dry cloth.
Repairs on valves, safety features and electric must only be carried out by an appointed CertoClav
partner or by the manufacturer. Use original CertoClav spare parts only.
Every two years the unit should be checked by an appointed retailer or CertoClav.

Spare parts

Art. no. 1300122
Art. no. 1310186
Gaskets should be replaced yearly. Long time cycles cause a higher wear and tear;
replacements could be necessary already after 6 months.
CertoClav Sterilizer GmbH
Georg Grinninger-Str. 37
A-4050 Traun / Austria
www.certoclav.com
Lid gasket
Service Set 2 (includes lid gasket)