Autoclave CertoClav MultiControl 2

Safety remarks

Use the autoclave chamber only with the MultiControl 2 Controller. Do not plug in directly to a
power supply. The chamber will heat in an uncontrolled way.
Never manipulate on safety devices. Operate only controller and chamber of the same series.
Do not mix lid and chamber of different types of autoclaves.
Surfaces are hot while in operation (metal and plastic parts). Wear protective gloves.
Use only in accordance with this manual. Store the manual in a space accessible to all users.
Must not be installed in a space that is subject to the danger of explosions
Use only with rated voltage.
Only for use in laboratories. Operate only by trained personnel. Do not use in domestic
settings.
Never operate without water. Target water amount: 1,25 liters (min. 1 liter)
Do not autoclave tightly sealed containers (especially glass bottles) since there is no
pressure compensation.
Use only water for steam creation.
Do not release steam in the direction of a person – danger of burning. We recommend to
connect a temperature resistant hose (inner diameter 10mm) to release hot steam into a
suitable container or sink. We offer this hose and container as accessory: Art. no. 8582012 –
hose 1.0m
If the pressure is released quickly when autoclaving liquids, there is a possibility that the
liquid will boil. I this should be avoided we recommend cooling the autoclave to 70°C before
opening the valve.
When running long time cycles danger of cooking dry and overheating. The autoclave could
be damaged. Check regularly water level.
Do not transport when filled. Do not open with force.
This autoclave must only be maintained by CertoClav or authorized retailers. Always use
original spare parts.