Autoclave - technical data

Overview

Technical data
Material
Dimensions
Volume
Chamber diameter
Chamber height
Useable height chamber 30 cm
Working data
Max. working temp
Max. working pressure
Rated pressure

Directives

Tested according:
2014/30/EU – EMC – Associated stan
o EN 61326-1:2013 Electrical Equipment for measurements, control and
laboratory use - EMC-Requirements Part 1: General requirements
o EN 61000-3-2:2006 + A1:2009 + A2:2009 Electromagnetic Compatibility (EMC)
— Part 3- 2: Limits - Limits for harmonic current emissions (Equipment input
current equal to or less than <= 16A per phase)
o EN 61000-3-3:2013 Electromagnetic Compatibility — Part 3- 3: Limits —
Limits for voltage fluctuation and flicker on the public low-voltage system
(equipment having an input current to or less <= 16 A per phase), and not
subject to conditional connection.
RoHS 2011/65/EU: Directive on the restriction of the use of certain hazardous
substances
EN 61010-1:2010
CertoClav Sterilizer GmbH
Georg Grinninger-Str. 37
A-4050 Traun / Austria
www.certoclav.com
Stainless steel
18 liters
26cm
35cm
140°C
2,7 Bar
4,1 Bar
support@certoclav.com
Certificates
Height
Width
Depth
Weight
Current
Power
dards:
Page 14 of 44
CE, TÜV
60 cm
44 cm
40 cm
10 kg
230V, 50/60 Hz
1300 Watt