Download Print this page

Jeio tech OV-11 Service Manual page 5

Advertisement

Classification
Symptom
Service Factor
Temperature
Temp doesn't increase
PCB Malfunction
1. Set the SV value higher than PV, Press the Start Button.
2. Check the HEAT LED is lightning on Display. (No Flickering, Lightning)
3. Check Terminal Resistance using of the Multi meter. (DC MODE)
If resistance doesn't measure, it is a malfunction.
Check point
- Replace the PCB.
1. "2-1-1. Side Cover panel disassembly" Reference
Solution
2. "2-1-3. Main board PCB disassembly" Reference
3. The assembly is reverse order of disjointing.
- 5 -

Advertisement

loading

This manual is also suitable for:

Ov-12