1-2. Power
Classification
Power
No Power
1. Check the Fuse using of the multi meter.
Check point
If Resistance is not measured, Fuse disconnection.
- Replace the Fuse.
Solution
1. "" Reference
2. The assembly is reverse order of disjointing.
Classification
Power
Replace the Fuse.
1. Measure output terminals No.2,5 When turning on the Main power.
If Power is not measured, Switch Malfunction
Check point
- Replace the Switch.
1. "2-1-1. Side Cover panel disassembly" Reference
Solution
2. Separate Switch wiring. (When assembling, location is a caution.)
3. Pull out the parts which are involved in the Switch upper & bottom from the panel.
4. Assembly is in reverse disassembly.
Symptom
Symptom
- 12 -
Service Factor
Fuse disconnection
Service Factor
Replace the Fuse.