Download Print this page

Jeio tech OV-11 Service Manual page 4

Advertisement

1. "2-1-1. Side Cover panel disassembly" Reference
2. "2-1-5. Relay disassembly" Reference
3. The assembly is reverse order of disjointing.
Classification
Symptom
Service Factor
Temperature
Temp doesn't increase
SSR Malfunction
1. Set the SV value higher than PV, Press the Start Button.
2. Check the HEAT LED is lightning on Display. (No Flickering, Lightning)
3. Check Terminal Resistance using of the Multi meter. (Resistance MODE)
If resistance doesn't measure, it is a malfunction.
Check point
- Replace the SSR.
1. "2-1-1. Side Cover panel disassembly" Reference
Solution
2. "2-1-4. SSR disassembly" Reference
3. The assembly is reverse order of disjointing.
- 4 -

Advertisement

loading

This manual is also suitable for:

Ov-12