1-6. Vacuum
Classification
Symptom
Vacuum
Vacuum leakage
* How to check the leaking point *
- Easily error parts. -
1. Check and replace Door packing with vacuum status.
2. Check if Vent Valve fixing bolt is loosen.(if loosens easily.)
3. Check Reducer's NUT on Temperature Controlling.
Check point
- If you still find any faulty on the above parts. -
1. Fill water in the chamber by 1 cm.
2. Let the chamber vacuum status.
3. Turn the chamber left/right ,and up/down to find air leaking.
4. If Chamber is OK, but pipes are leaking, replace the leaking pipe.