Classification
Power
Power shut off during operation.
1. Check the fuse.
2. If fuse is down, replace it..
Check point
3. Check Resistance both ground wire and Main plug using of the multi meter.
If resistance is measured, it is Wiring insulation malfunction.
Solution
- Replace the wire. ("2-2-9. Wiring short circuit Inspection" Reference)
Classification
Power
Power shut off after turning on
1. Check the fuse.
2. Check an electric outlet capacity connected the unit.
Check point
3. Check total units connected the multi electric outlet.
If total units connected the multi electric outlet is over than the outlet capacity, multi electric outlet malfunction.
Solution
- Use the multi electric outlet under capacity.
Symptom
Symptom
- 17 -
Service Factor
Wiring insulation malfunction
Service Factor
Electric capacity over in Multi outlet