Jeio tech OV-11 Service Manual page 16

Advertisement

Classification
Power
Power shut off during operation.
1. Check an electric outlet capacity connected the unit.
Check point
2. Check total units connected the multi electric outlet.
If total units connected the multi electric outlet is over than the outlet capacity, multi electric outlet malfunction.
Solution
- Use the multi electric outlet under capacity.
Classification
Power
Power shut off after turning on
1. Check the fuse.
2. Separate the heater terminal and wire.
3. Measure Resistance both the heater terminal and heater exterior.
Check point
If fuse is down and heater resistance is measured, Heater leakage.
- Replace the heater.
1. "2-1-8. Heater disassembly" Reference
Solution
(When replacing a Heater, Inner Chamber Jacket, Chamber Disassembly are very complicated, Do not replace the heater in the field.)
2. The assembly is reverse order of disjointing.(Heater should be checked resistance for inferiority.)
Symptom
Symptom
Electric capacity over in Multi outlet
Heater leakage
- 16 -
Service Factor
Service Factor

Advertisement

loading