1. Service
1-1. Temperature
Classification
Temperature
Temp doesn't increase
1. After setting 100℃ then press Start Button to operate the unit.
2. Check the Heater temp itself. (Be careful heater is very hot.)
3. Measure resistance checking positive terminal of heater using of the multi meter.
Check point
If resistance doesn't measure, it is a malfunction
- Replace a heater.
1. "2-1-8. Heater disassembly" Reference
Solution
(When replacing a Heater, Inner Chamber Jacket, Chamber Disassembly are very complicated, Do not replace the heater in the field.)
2. The assembly is reverse order of disjointing.(Heater should be checked resistance for inferiority.)
Classification
Temperature
Temp doesn't increase
1. 1. Press the Start Button to operate the unit, Measure Voltage of Output terminal using of the
multi meter.
If Voltage doesn't measure, it is a malfunction
Check point
Solution
- Replace the Relay.
Symptom
Symptom
- 3 -
Service Factor
Heater Malfunction
Service Factor
Relay Malfunction