Option CR3, or CRP, or ALV Installation Kit
Manually verify operation
Option CR3, Second IF Output
1. Preset X-Series analyzer. Assure mode is Spectrum Analyzer.
2. Display the 50 MHz calibrator signal. Press
, key in 50 MHz. Press
FREQ/Channel
display at -25 dBm.
3. Enable Option CR3. Press
Out
.
4. Connect a spectrum analyzer to the rear panel Aux IF Out port using test cable and appropriate
adapters.
Spectrum analyzer settings:
Center frequency: 322.5 MHz
Span: 1 MHz
Bandwidth: 100 kHz
The spectrum analyzer should display a −33 dBm ± 5 dB signal
Option CRP, Arbitrary IF Output
1. Preset X-Series analyzer. Assure mode is Spectrum Analyzer.
2. Display the 50 MHz calibrator signal. Press
, key in 50 MHz. Press
FREQ/Channel
display at -25 dBm.
3. Enable Option CRP. Press
Set the Arbitrary IF Out frequency to 70 MHz using the key pad.
4. Connect a spectrum analyzer to the rear panel Aux IF Out port using test cable and appropriate
adapters.
Spectrum analyzer settings:
Center frequency: 70 MHz
Span: 1 MHz
Bandwidth: 100 kHz
The spectrum analyzer should display a −33 dBm ± 5 dB signal
8
Input/Output
,
Span
Zero Span
,
,
Input/Output
More
Output Configuration
Input/Output
,
Span
Zero Span
,
,
Input/Output
More
Output Configuration
,
RF Calibrator
,
50 MHz
. You should see a flat line across the
,
Aux IF Out
,
RF Calibrator
,
50 MHz
. You should see a flat line across the
,
Aux IF Out
Installation Note
. Press
,
Second IF
. Press
,
.
Arbitrary IF
N9030-90036