Alternate Installation Procedure
The License File can be manually installed over USB or LAN by placing the license file in the
following folder on the signal analyzer
C:\Program Files\Agilent\licensing
Verify the Installation
1. Cycle the power on the signal analyzer.
2. Press
,
System
Show
3. Verify that the newly installed option appears on the list.
Functional Check:
Within one year of the initial calibration date of the instrument this option is fully calibrated with no
further adjustment or verification testing. To ensure that this newly installed option has been enabled
correctly the following functional check is recommended.
1. Press System, Show, Errors, Clear Error Queue. This will remove out-of-date errors.
2. Press
System
,
Alignments
3. Watch for any errors during the instrument alignment.
4. Press
,
System
Show
5. Verify that there were no errors during the alignment.
Installation Note
N9030-90036
,
to display a list of installed options.
System
,
Align Now, All
.
,
to view the instrument error log.
Errors
Option CR3, or CRP, or ALV Installation Kit
7