Photodiode test using femto/picoammeter and electrometer/high resistance meter
Hide thumbs Also See for B2980A Series:

Advertisement

17 | Keysight | Photodiode Test Using the Keysight B2980A Series - Technical Overview
Set up the B2911A via front panel operation
1. Set the Source Function to Current
a. Press
repeatedly until Single View for Channel 1 is shown.
b. Press
and then press
c. Press
and then set the Channel 1 Source Value to 0 A.
2. Set the Voltage Limit Value
a. Press
and then set the Channel 1 Limit Value to 2.5 V as shown.
3. Set the Aperture Time
a. Press
,
and then press
(0.01 PLC).
4. Enable remote sensing
a. Press
,
b. Press
to select Sensing Type and then press
c. Then press
to set the Channel 1 Source Function to Current.
to set the Aperture Time to SHORT
,
to open the Output Connection dialog.
to apply the changes.
to enable remote sensing.