16 | Keysight | Photodiode Test Using the Keysight B2980A Series - Technical Overview
Setting up the B2985A/87A from its front panel (continued)
6. Enable the Voltage Source and Ammeter.
a. Press Voltage Source
b. Press Ammeter
Controlling the B2985A/87A using SCPI commands
If you prefer to control the B2985A/87A remotely using SCPI remote commands, then
the following material explains how to do this.
The series of commands shown in group "A" configure measurement parameters such
as measurement range and aperture time. Next the series of commands shown in group
"B" should be sent to set up the voltage source. Finally, the series of commands shown in
group "C" should be sent to set up the B2985A/87A's trigger parameters.
to enable the Voltage Source.
to enable the Ammeter.