Table power supply (external):
Rated voltage
Frequency
Input power max.
Calorimeter:
Rated voltage
Input power max.
Fuses (internal)
Perm. on-time
Protection class as per DIN EN 60 529
Protection class
Overvoltage category
Contamination level
Perm. ambient temperature
Perm. ambient humidity
Operation at a terrestrial altitude
Dimensions
Weight
Measurement ranges
Measuring mode / measurement times
Interfaces
Subject to technical changes!

Technical data

Hz
W
W
°C
%
mm
kg
J
min
100 - 240 V AC
50 / 60
150
24 V DC 5A
150
1x 2,5 AT
Continuous operation
IP 20
III
2
II
20 ... 25 (constant)
80
2000 m above sea level
400 x 400 x 400 (B x T x H)
21
40.000
approx. 17 /Isoperibol
approx. 8 /Dynamic
14 /Time control
approx. 17 /Manual
1 x parallel (Centronics)
1 x serial (RS 232)
30