Pump
Switch the pump on with this menu item. The
outer vessel will be filled and rinsed (see also sec-
tion 4.7).
Requirement:
Check that there is water in the tank.
Empty outer vessel
This menu item allows you to empty the outer
vessel.
Note:
Shut-down, completely empty appliance.
Requirement:
The emptying hose must be locked in place in
plug-in coupling (section 4.2).
Cleaning and maintenance
In order to ensure trouble-free operation over a
long period of time, it is important to carry out
the following maintenance work to the calorime-
ter system.
Disconnect main plug prior
to cleaning!
8.1 Inner vessel filler
Remove the filter element from the inner vessel
and clean the inner vessel filter in clear water or
an ultrasonic bath if there is any dirt visible. Also
clean the inner vessel, by re-moving the positio-
ning mount for the decomposition vessel. Place
the filter back on the inner vessel connector after
cleaning.
Operation without the filter element causes dirt
in the valves and thus appliance failure.
8.2 Filler
Reset
This menu item allows you to restore the default
settings.
Reinit
This menu item allows you to reinitialise the
A / D converter.
There is also a filter in the filler. This prevents dirt
in the storage container. If the filter is dirty, remo-
ve the entire filler and clean the filter (see section
4.1, ).
8.3 Micro filter
In the back plate of the equipment is additionally
a micro filter. If the micro filter is dirty, the water
flow in the equipment is reduced. Thus the water
filling time of the inner vessel extends. Thereby
if the upper time limiting is exceeded by 120 se-
conds, the system indicates the disturbance and
breaks the measurement off.