Guidance and manufacturer's declaration – electromagnetic immunity
The InMode is intended for use in the electromagnetic environment specified below. The customer or the user
of the InMode should assure that it is used in such an environment.
IMMUNITY
IEC 60601
test
TEST LEVEL
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to
80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2,5 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which
the InMode is used exceeds the applicable RF compliance level above, the InMode should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be necessary,
such as re-orienting or relocating the InMode.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
InMode BodyFX Operator Manual
Electromagnetic environment – guidance
Compliance
level
[3] V
Portable and mobile RF communications equipment
should be used no closer to any part of the InMode,
including cables, than the recommended separation
distance calculated from the equation applicable to the
[3] V/m
frequency of the transmitter.
Recommended separation distance
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
be less than the compliance level in each frequency
b
range.
Interference may occur in the vicinity of equipment
marked with the following symbol:
a