Section 8 – Troubleshooting
The InMode System provides monitoring of all critical parameters to ensure safety of
patient and user. If any of the following faults are detected system automatically goes to
STAND BY mode.
Description of Faults with All Handpieces
System did not turn on
Check power cord connection
Check that main switch on rear panel is on
Check fuses on back panel of the System
Call Technical Service if problem persists
Software plug missing
The software plug is not inserted
Checksum
The software was not loaded properly from software plug
Check plug connection and reboot the System
Call Technical Service if problem persists
Fault H8002 - Handpiece is not connected
Check connection of Handpiece
Replace Handpiece
Call Technical Service if problem persists
Fault H8005 – System Memory Fault
Call Technical Service if problem persists
Fault H8601 – Distributor Card Connection Fault
Call Technical Service if problem persists
Fault H8609 – Water Temperature Fault
Call Technical Service if problem persists
Fault H860B – Water Flow Fault
Call Technical Service if problem persists
Laser Related Faults - H8003, H8006, H8007
Call Technical Service if problem persists
InMode BodyFX Operator Manual
35