Section 3 - System Installation
Electrical Requirements
The System will require a separate line supply of single phase (100Vac; 15A) or
(115Vac; 15A) or (230Vac; 15A) or (240Vac; 15A) 50-60Hz.
Power receptacles must be within 15 feet of the System site.
The System should not share a power line with other equipment.
Power receptacle must include protective earth, and must be checked before
connecting the system.
For continued protection against fire, replace the fuse only with
one of the same type and rating.
Proper grounding is essential for safe operation.
Environmental Requirements
Corrosive materials can damage electronic parts; therefore, the System should
operate in a non-corrosive atmosphere.
For optimal operation of the System, maintain room temperature between
20º-27ºC (68º-79ºF) and relative humidity of less than 80%.
Equipment List
The System includes the following:
System platform
BodyFX Handpiece
BodyFX Handpiece cradle
Foot switch
Operator manual
Power cord
InMode BodyFX Operator Manual
15