Appendix 1 - Main Parameters - Kane 457 Manual

Flue gas and ambient air analyser with direct co2 measurement
Hide thumbs Also See for 457:
Table of Contents

Advertisement

APPENDIX 1 - MAIN PARAMETER:
Here are the legends used and what they mean:
O
:
Oxygen (Calculated) reading in percentage (%)
2
COp :
Carbon monoxide (Measured) reading displayed in ppm (parts per
million). If '- - - -' is displayed there is a fault with the CO sensor or the
instrument has not set to zero correctly. Switch off instrument and try
again.
COv:
Averaged ambient CO reading in ppm.
COx:
Maximum ambient CO reading in ppm
CO
:
Carbon dioxide (measured) reading in percentage (%).
2
CO
p:
Carbon dioxide (measured) reading in parts per million (ppm)
2
CO
v:
Averaged Carbon dioxide reading in ppm
2
CO
X:
Maximum ambient CO
2
TF :
Temperature measured by the flue gas probe in centigrade (
displays '- OC -' if the flue probe is disconnected or faulty.
TI :
If an inlet temperature probe (optional) is connected into the T2 socket
during its' countdown, the measured temperature from the inlet probe
will be used as the inlet temperature.
If an inlet temperature probe is not connected to the analyser during
countdown the measured temperature from the flue probe will be used
as the inlet temperature.
If neither probe is connected during countdown the analyser's internal
ambient temperature will be used as the inlet temperature.
T Nett :
Nett temperature calculated by deducting the INLET temperature from
the measured FLUE temperature. It displays '- OC -' if the flue probe
is not connected or broken.
EfC :
Combustion efficiency calculation displayed in percentage either as
Gross Ef(G) or Nett Ef(N) or Condensing Nett Ef(C) - Use MENU to
change. The calculation is determined by fuel type and uses the
calculation in British Standard BS845. The efficiency is displayed
during a combustion test, '- - - -' is displayed while in fresh air.
Loss :
Losses calculated from oxygen and type of fuel. Displays reading
during a combustion test. '- - - -' is displayed while in fresh air.
reading in ppm
2
o
C). It
Page 53

Hide quick links:

Advertisement

Table of Contents
loading