Adjust Thecornererror; Corner Error Inspection - KERN ADJ Series Service Manual

Analytical balance
Hide thumbs Also See for ADJ Series:
Table of Contents

Advertisement

If the balance displays CAL NO, this indicates the balance is not yet correctly calibrated.
Please press ON/OFF key to reboot the balance and re-calibrate.
There are several reasons why the balance could not be calibrated:
a. Unstable displayed value of the balance (faulty sensor or Main PCB or interference between
the sensor and its peripheral devices).
b. The count value of the balance falls outside the specification range.
c. Inaccurate calibration weight or deviation.

11 Adjust theCornerError

11.1 Corner Error Inspection

Before checking the corner error of the balance, please adjust the level first.
ADJ-SH-e
1910
The screen shows CAL..... After a few seconds, it
displays CAL UP
Remove the weight and the screen displays
CAL END
The calibration is over and the balance displays
0.0000g
Put a 100 g weight (half the capacity of
the balance) on the center of the
weighing pan, read and record the
displayed value.
Move the weight to around 2/3diameterof
the weighing pan, and record four
displayed values when the balance is
31