10.3.1 Factory External Weight Calibration

Before calibration, please make sure you have preheated the balance for
30 minutes and adjusted the level.
28
Add a 50g (class E2) weight on the center of the
weighing pan. 150g in total.
After a few seconds the screen shows 0.0000g.
Linearity calibration is completed and the balance
returns to normal weighing mode.