Chapter 4 Edit a plate before starting a run
4
Options to edit the method of a plate file

Set up or remove the Auto Delta feature

58
2. Tap the temperature field in the appropriate VeriFlex
3. Enter a temperature, then tap Enter.
The temperatures in adjacent zones must have a difference of ≤5°C.
4. Tap Done.
A stage with VeriFlex
5. In the Method tab, tap
Note: The VeriFlex
Indicates that this is the Method tab where the VeriFlex
1
Indicates that VeriFlex
2
Note: The Auto Delta feature is only available in the PCR stage.
1. In the Method tab, tap Edit4Advanced4VeriFlex
Note: Use
or
2. Tap
(Edit) in one of the PCR stages.
3. Tap Auto Delta settings.
See Figure 8 on page 57.
4. Enter the starting cycle to apply Auto Delta settings.
5. Set the Delta temperature slider to the
enter the numerical value for the difference in the temperature.
6. Set the Delta time slider to the
numerical value for the difference in the time.
7. Tap Done.
A PCR step with an Auto Delta setting applied to it is denoted with an
Zones is denoted with a
4Remove4Done to remove the VeriFlex
Zones setting is removed from the Method tab.
Zones are assigned. Tap to remove the VeriFlex
to navigate through the method if it is not all displayed.
position or the
QuantStudio
6 Pro and 7 Pro Real-Time PCR Systems User Guide
Zone.
icon.
1
2
Zones settings can be removed
Zones.
Zones / Auto Delta.
position or the
position, then
position, then enter the
Zones.
.