CONTENTS Introduction Clinical Applications Assembly Installation and Disassembly Operation Cleaning Maintenance/ Troubleshooting Accessories Warranty and Service Technical Specification Parts List Warnings, Cautions & Notes Warnings given in this manual identify possible hazards in procedures or conditions, which if not correctly followed, could result in death, injury or other serious adverse reactions Cautions given in this manual identify possible hazards in procedures or conditions, which if not correctly followed, could result in equipment damage or failure.
1. Introduction This manual provides instructions for the use and maintenance of the ArjoHuntleigh Minuet ® bed with model numbers listed in the table below MODEL NUMBER 100500AUS 100520AUS DESIGN FEATURE Four section mattress platform Two section mattress platform...
3. Assembly Check that you have the following parts: Head end assembly Foot end assembly Head end mattress platform assembly Foot end mattress platform assembly Control handset Six clamp screws (loose) and Allen key Make sure you have enough room to turn the bed from its side to upright when assembled.
Page 5
3. Stand the assembled components on one side with the connector sockets on the backrest actuator or control box (A) uppermost. Assemble the head end deck assembly to the foot end deck assembly. Assemble the foot end to the deck. AH-MIN V3 08/10/17...
Page 6
There are a total of 6 clamp screws provided for assembling the bed. Assemble the first 3 of these clamp screws (B) on the underside of the side rail tube of one length of the bed. Make sure all joints are fully inserted and tighten all 3 clamp screws securely using supplied Allen key.
Page 7
Carefully thread the foot end actuator cable through the support loop (G) on the deck frame and connect it to socket 4 on the backrest actuator/control box. Adjust the cable coils in the loop so the cable is under tension but not strained.. Connect the control handset cable to socket 1 on the backrest actuator or control box.
Page 8
Do not expose the equipment, especially the mattress, to naked flames such as cigarettes etc. Use only ArjoHuntleigh accessories designed or approved for use with the bed. o When moving bed, wrap electrical power cable around Cable Tidy, located at bed head end in order to avoid damage to cable.
Page 9
Power Operation When a handset push button is pressed, the appropriate bed function will operate until the button is released. WARNING Some functions may cause injury to the patient if unintentional movement occurs. WARNING If the Handset is dropped to the floor, caught between bed parts or sustains any damage or suspected damage in any way, immediately stop using the Handset and perform the full Handset Checks as detailed in Section 7 - Maintenance.
The following preventative maintenance procedures should be carried out at intervals of 12 months or as specified. All service and repair activities must only be performed by properly qualified and trained technicians approved by ArjoHuntleigh. General Examine the bed for obvious signs of damage. Make sure that all nuts, bolts, clamp screws and other fastenings are tight and not missing.
Page 11
3 below. Preventative Maintenance Checks ArjoHuntleigh recommends that the full check and testing regime is conducted between each bed use or worst case during annual Preventative maintenance. This should take no more than one (1) minute to complete.
Page 12
travels further when depressed Immediately quarantine the bed and report the damage to maintenance staff or replace the Handset, otherwise contact an ArjoHuntleigh approved Technician. The following table identifies some fault symptoms, their possible cause and suggest some simple checks and corrective actions. If these actions fail to solve the problem consult the ArjoHuntleigh Service Department on Freecall 1800 072 040.
The maximum safe working load of the self help pole is 76.5kg. The pole must be used when the handle is central above the mattress platform A full range of accessories are available from ArjoHuntleigh. Details can be provide upon request.
. Technical Specifications Dimensions Overall Length 2200mm Overall Width 920mm Height to Mattress Platform 230mm 630mm Backrest Angles To 70º Leg Elevation: 100500 AUS 18º 100520 AUS 15º Tilt: 100500AUS / 100520AUS 10 Head up to 10 Head down Mattress Size: 1980 x 860 x 150-170mm Product Weight (approximate) Foot End Assembly (standard)
Page 16
WEBSITE www.arjohuntleigh.com PERTH—ADELAIDE—MELBOURNE—SYDNEY—BRISBANE As our policy is one of continuous improvement, we reserve the right to modify designs without prior notice. ArjoHuntleigh (ABN 75 007 200 900) A member of the Huntleigh Technology PLC Group of Companies AH-MIN V3 08/10/17...
Need help?
Do you have a question about the Minuet 100500AUS and is the answer not in the manual?