Cleaning Air Filters

The Aixplorer® ultrasound system is equipped with air filters. These
should be inspected every week and cleaned as needed. At first switch
on of the system after the first day of the month, the system displays an
icon as a reminder to check the air filters. Please refer to the following
air filter cleaning procedure, and click
The air filters are located:
• in a slot at the rear side of the system
• in a slot at the bottom side of the system.
To remove the rear air filter, open the cover on the back. To remove the
side air filter, open the left side cover with a pen. It is recommended
that air filters be cleaned monthly or more frequently if a significant
accumulation of dust or debris is noted.
NOTE
Please note that side filter may be missing on some models that do not
require it anymore.
Air filters may be removed, washed with mild soap and rinsed
thoroughly with water. Allow the filters to air dry before re-installing in
the system. Spare air filters are provided for immediate use. Additional
air filters can be ordered from SuperSonic Imagine Customer Service.
6
OK
on the pop-up window.