pH VALUES AT VARIOUS TEMPERATURE
Temperature has an effect on the pH. The calibration buffer solutions
are affected by temperature changes to a lesser degree than normal
solutions.
Please refer to the following chart to perform the pH calibration:
TEMP
ºC
ºF
0
32
5
41
10
50
15
59
20
68
25
77
30
86
35
95
40
104
45
113
50
122
55
131
60
140
65
149
70
158
For instance, if the buffer temperature is 25 °C (77 °F), calibrate to
read on the display pH4.01 or 7.01 or 10.01. If the buffer temperature
is 20 °C, calibrate to read on the display pH4.00 or 7.03 or 10.06.
If the buffer temperature is 50 °C, calibrate to read on the display
pH4.06 or 6.98 or 9.82.
pH BUFFERS
4.01
6.86
7.01
4.01
6.98
7.13
4.00
6.95
7.10
4.00
6.92
7.07
4.00
6.90
7.05
4.00
6.88
7.03
4.01
6.86
7.01
4.02
6.85
7.00
4.03
6.84
6.99
4.04
6.84
6.98
4.05
6.83
6.98
4.06
6.83
6.98
4.08
6.84
6.98
4.09
6.84
6.98
4.11
6.84
6.99
4.12
6.85
6.99
30
9.18
10.01
9.46
10.32
9.39
10.24
9.33
10.18
9.27
10.12
9.22
10.06
9.18
10.01
9.14
9.96
9.11
9.92
9.07
9.88
9.04
9.85
9.01
9.82
8.99
9.79
8.97
9.77
8.95
9.76
8.93
9.75