Download Print this page

Keysight Technologies PNA-X N5241B Installation Note page 29

To upgrade option 217 to option 417 for analyzers with serial numbers prefixed my/sg/us5201 and above
Hide thumbs Also See for PNA-X N5241B:

Advertisement

Description of the Upgrade
Installation Procedure for the Upgrade
Step 17.Install the Test Set Cables
Follow instructions carefully when making cable connections, especially
wire harness connections. Incorrect connections can destroy components,
resulting in additional customer costs.
Be careful not to damage the center pins of the semi-rigid cables. Some
flexing of the cables may be necessary but do not over-bend them.
Use a 5/16-in torque wrench set to 10 in-lbs on all cable connections
except the front and rear panel cable connectors. Torque these
connections to 21 in-lb.
Cables that are to be reinstalled are designated with "reuse." Part numbers
in a normal weight font are for analyzers with serial numbers
MY/SG/US5310 and below. Part numbers in a bold font are for analyzers
with serial numbers MY/SG/US5321 and above.
1. Install the following semi-rigid cables in the order listed. To see images
Installation Note N5242-90122
showing the location of these cables, click the Chapter 6 bookmark
"Bottom RF Cables, 4-Port, Option 417" in the PDF Service Guide
parts are listed in
Table 1 on page
For all analyzer serial numbers:
— W26 (N5222-20034) A32 port 2 coupler to front-panel Port 2 CPLR
ARM
— W93 (reuse) (N5242-20047 or N5222-20037) Port 2 RCVR B IN to
A45 port 2 receiver attenuator
— W84 (reuse) (N5242-20046 or N5222-20036) A37 port 2 source
attenuator to front-panel Port 2 SOURCE OUT
— W135 (reuse) (Opt 029A only) (N5242-20073 or N5242-20278)
Front-panel Port 2 RCVR B IN to A45 port 2 receiver attenuator
— W128 (reuse) (Opt 029A only) (N5242-20134 or N5242-20303) A53
port 2 bypass switch to front-panel Port 2 SOURCE OUT
— W127 (reuse) (Opt 029A only) (N5242-20116 or N5242-20292) A37
port 2 source attenuator to A53 port 2 bypass switch
— W130 (reuse) (Opt 029A only) (N5242-20133) A53 port 2 bypass
switch to A54 port 2 bridge
— W129 (reuse) (Opt 029A only) (N5242-20117 or N5242-20293) A53
port 2 bypass switch to A54 port 2 bridge
1. See
"Downloading the Online PNA Service Guide" on page
9.