Battery Removal And Replacement - Polycom RealPresence Practitioner Cart 8000 User Manual

Table of Contents

Advertisement

RealPresence
BATTERY REMOVAL AND
REPLACEMENT
Only use batteries specified by
Rubbermaid Medical. Failure to do so
will void power system warranty. Call
customer service for details.
Do not replace battery in oxygen rich
environments as it is possible for
sparking to occur.
Always turn off the cart and unplug the
power cord from the wall outlet when
removing the battery.
There are two conditions when a battery should be
disconnected:
1.
When the battery needs replacing
2.
When the cart will be placed into long-
term storage (more than 60 days without
use)
TOOLS REQUIRED
1.
Ball End Metric Tool Set
2.
10mm Open End Wrench
STEPS FOR BATTERY REMOVAL
1.
Power Down the System
a.
Turn off any components
plugged into cart (example –
PC, Monitor)
b.
Turn off Power System by
Holding On/Off switch for 2
seconds
LONG TERM BATTERY STORAGE (Greater than
2 Months)
WARNING: If cart will not be used for
more than 2 months, turn off the cart by holding
the on/off button. The battery will need to be
recharged every 30 days to prevent damage.
If you wish to store the battery separate
from the cart, store battery in a cool dry location.
UNPLUG POWER CORD FROM THE OUTLET
Manufactured by: Rubbermaid Healthcare • 16905 Northcross Drive • Suite 200 • Huntersville NC 28078
Polycom RealPresence® Practitioner Cart® 8000 User Manual p/n 1793149 Rev C 03/13
®
®
Practitioner Cart
8000
14
REMOVE BATTERY ACCESS PANEL AND
BATTERY ACCESS STRAP
a.
Unplug power cord from battery
access panel
b.
Using 4mm hex key or 5/32" remove
two fasteners on lower rear panel of
battery access panel and place aside
c.
Remove Battery Retainer Strap by
depressing the tab on the clasp and
unthreading strap from buckle
a
DISCONNECT BATTERY CABLES FROM THE
BATTERY
a.
Remove the black boot from the
terminal.
b.
Use wrench or screwdriver to
remove bolt from battery terminal.
WARNING: Do not allow any metal
objects or Tools to contact both the
terminal and other metal parts
(including the other battery terminal)
at the same time.
c.
Remove the red terminal cover
d.
Use wrench or screwdriver to
remove bolt from the battery terminal
e.
Place disconnected cables off to side
f.
Grasp handles on battery and slowly
pull out
a
b
c

Advertisement

Table of Contents
loading

Table of Contents