Guidance and manufacturer's declaration- electromagnetic immunity
The CMCT is intended for use in the electromagnetic environment specified below. The customer or
the user of the CMCT should assure that it is used in such an environment.
Immunity Test
Radiated Immunity
IEC 61000-4-3
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
CMCT Monitoring System
Instructions for Use
IEC 60601 Test
Compliance
Level
Level
10 V/m,
10 V/m
frequency
80MHz-2500
MHz
9-28 V/m
358-5800MHz
PM
18Hz, 217Hz
Electromagnetic environment- guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the ZOLL CMCT including
cables, than the recommended separation
distance calculated from the equation
appropriate to the frequency of the
transmitter.
Recommend separation distance
d = 1.2 √P
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2.5 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the recommended
separation distance in meters (m).
Field strengths from fixed RF transmitters
as determined by an electromagnetic site
a
survey,
should be less than the
compliance level in each frequency
b
range.
Interference may occur in the vicinity of