ZOLL CMCT Monitoring System Instructions For Use Manual page 29

Table of Contents

Advertisement

Guidance and manufacturer's declaration- electromagnetic immunity
The CMCT is intended for use in the electromagnetic environment specified below. The customer or
the user of the CMCT should assure that it is used in such an environment.
Immunity Test
Electrostatic discharge
IEC 61000-4-2
Immunity to Magnetic Field
IEC 61000-4-8
Electrical fast
transient/burst
IEC 61000-4-4
Conductive Surges
IEC 61000-4-5
Conducted Disturbances
IEC 61000-4-6
Voltage Dips and Short
Interruptions
IEC 61000-4-11
CMCT Monitoring System
Instructions for Use
IEC 60601
Compliance
Test Level
Level
±8 kV contact
±8 kV contact
±15 kV air
±15 kV air
30 A/m ,
30 A/m
50Hz/60Hz
±2 kV, 100 kHz
±2 kV
Repetition
frequency
±1 kV line to
±0.5, ±1
line
3V frequency
0.15MHz-
80MHz
6V in ISM and
amateur radio
bands between
0.15MHz and
80MHz
0 % U T; 0,5
0 % U T; 0,5
cycle
cycle
At 0 ° , 45 ° , 90
At 0°, 45°, 90°,
° , 135 ° , 180
135°, 180°,
225°, 270° and
° , 225 ° , 270
315°.0 % U T; 1
° and 315 °
cycle
0 % U T; 1 cycle
and
and
70 % U T; 25/30
70 % U T; 25/30
cycles
cycles
Single phase:
Single phase: at
at 0°
0 °
Electromagnetic environment-
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
3V
Mains power quality should be that
of a typical commercial or hospital
environment.
6V
Mains power quality should be that
of a typical commercial or hospital
environment.
If the user of the device requires
continued operation during power
mains interruptions, it is
recommended that the device be
powered from an uninterruptible
power source.
Page 29 of 31

Advertisement

Table of Contents
loading