Advertisement

Configuration
September 2019
14.
15. Attach a length of calibration tubing (PVC tubing, 3/16-in. ID, 5/16-in. OD) from the
54
WARNING
Toxic gas
Before performing the next step, verify that the regulator is closed to avoid
releasing target gas into the air during calibration.
Install a regulator on the target gas source.
regulator on the target gas source to the fitting on the IP filter assembly (part
number 00628-9000-0001).
Reference Manual
00809-0100-4928
Emerson.com/Rosemount