11. Select OK to accept the new zero.
12. On the Calibrate Sensor screen, enter a gas concentration level that corresponds to
13. Select OK.
Rosemount 928
the concentration of calibration gas that will be applied during calibration.
For oxygen, use 20.9 percent oxygen from clean air. This step may be performed
with surrounding air if no contaminants are present.
Configuration
September 2019
53