14.0 ERROR CODES

During the initial power-on testing or during operation, the scale may show an error
message. The meaning of the error messages is described below.
If an error message is shown, repeat the step that caused the message. If the error message
is still shown then contact your dealer for support.
ERROR
DESCRIPTION
CODE
Err 4
Initial Zero is greater than
allowed (4% of maximum
capacity) when power is
turned on or when [Zero] is
pressed.
Err 6
A/D count is not correct
when turning the scale on.
Err 8
High limit input error
Err 9
Low limit input error
FAIL H or
Calibration error
FAIL L
© Adam Equipment Company 2018
POSSIBLE CAUSES
Weight on the pan when turning the scale
on.
Excessive weight on the pan when zeroing
the scale.
Platform is not installed.
Improper calibration of the scale.
Damaged load cell.
Damaged Electronics.
Load cell is damaged.
Electronics is damaged.
Low limit is set first, then the high limit is set
lower than the low limit and high limit not
equal to zero.
High limit is set first, then the low limit is set
higher than the high limit and low limit not
equal to zero.
Improper calibration (should be within +10%
of the factory calibration). The old calibration
data will be retained until the calibration
process is complete.