Mch-1000(I) Control Module Check - CardioQuip MCH-1000 Operators & Service Manual

Cardiopulmonary bypass temperature controller
Hide thumbs Also See for MCH-1000:
Table of Contents

Advertisement

CardioQuip Modular Cooler-Heater Operator/Service Manual
WARNING: Proper surgical procedures and techniques are the responsibility of
!
the operator.
WARNING: To avoid the risk of electrical shock, this equipment must only be
!
connected to a supply mains with protective earth.
WARNING: Use a dedicated 20 Ampere AC "hospital grade" grounded outlet.
!
WARNING: This device must not be connected to a heat exchanger or blanket
!
with a manufacturer recommended max. pressure rating of less than 83 kPa (<12
psi). Rupture or leakage may occur in the heat exchanger resulting in patient injury
and/or damage to the system.
WARNING: During clinical operations, do not drain equipment connected to the
!
MCH. Draining attached equipment will drain the entire heat exchanger network.
WARNING: Do not operate the MCH with blocked water flow. Hoses must be
!
attached, and any valves must be open, so that water can flow freely. Operating the
MCH without hoses, or with closed valves on the hoses, will damage the pump and
void the warranty.
CAUTION: Position the MCH to allow easy access to the mains power plug. The
!
operator must be able to quickly unplug the device from mains power if necessary.
CAUTION: Do not completely drain the system while in an operating mode. Dry
!
operation will cause equipment damage. Use the tank drain to remove water from the
tank for adding additional ice, if necessary.
CAUTION: Do not use sharp objects to adjust or select switches on the control
!
panel.
CAUTION: Do not overtighten the control panel clamp or articulated arm lock.
!
Mechanical damage may result.

MCH-1000(i) Control Module Check

1. Unpack the Control Module and visually inspect for shipping damage
(scratched or cracked screen, dents, broken or exposed wires, bent pins,
etc.).
2. Using the articulated arm, mount the Control Module on one of the
convenience handles on the MCH base unit, or on a suitable pole within
reach of the MCH base unit with the Control Module's 15ft (3m) cable.
3. Plug the Control Module connector into the front panel of the MCH base unit.
Page 33