CardioQuip Modular Cooler-Heater Operator/Service Manual
Defrost
The MCH incorporates a built-in defrost cycle,
which utilizes the heaters in a special recirculating
mode to melt the ice in the cold water tank. Once
the water in the tank reaches room temperature,
the MCH automatically returns to STANDBY
mode. The DEFROST function is accessed from
the SETTINGS screen. To defrost properly, the
MCH must either have a hose loop connected, or
have its recirculation valve (if so equipped) set to
INTERNAL mode.

Run

The RUN screen may be accessed from the TESTING or STANDBY screens by touching the
"Run" button at the bottom of the screen. In this mode, the pump is running, water is circulating,
and temperature control is active.
System mode indicator
Current temp display
Operational status
Operational indicators
CPG Module controls
Control buttons
The temperature
control status shows
the current system
function
(COOLING,
HEATING, AT
TEMP, DEFROST)
!
Customizable message
Temperature control
buttons
Low Ice indicator
Pump speed control
button
The pump speed
control button shows
the current pump speed.
When priming or
defrosting, or heating
when Auto Pump High
is enabled, pump speed
is forced high, and is
indicated by this
"OVERRIDE"
message.
Page 17