Standby

In STANDBY mode, the pump and heaters are off, and the operator is able to preset the desired
temperature and other settings in preparation for entering RUN mode.
System mode indicator
Current temperature display
Operational indicators
CPG option controls
Control buttons
To set the desired temperature in tenth-degree
increments, touch the "+" or "–" temperature control
button. Holding the button for more than one second
will cause the temperature setting to change more
rapidly. Alternatively, touching the "Set Point"
temperature field will cause a numeric keypad to
appear onscreen, allowing direct input of the desired
temperature. After entering the desired temperature,
touch the "ENT" button to confirm, or touch "ESC"
to cancel.
The temperature setting, temperature presets, and customized message are stored in non-volatile
memory, and will be retained until changed.
CardioQuip Modular Cooler-Heater Operator/Service Manual
To modify the customizable message at the top right
of the screen, touch the message field to bring up an
alphanumeric keypad onscreen.
After entering the desired customized message (up to
12 characters; e.g., "CARDIOPLEGIA"), touch the
"ENT" button to confirm, or touch "ESC" to cancel.
Customizable text
Temperature control
buttons
Pump speed control