Mch Operation; Unpacking; Physical Inspection; Initial Operation Check - CardioQuip MCH-1000 Operators & Service Manual

Cardiopulmonary bypass temperature controller
Hide thumbs Also See for MCH-1000:
Table of Contents

Advertisement

Page
32

MCH Operation

Unpacking

Carefully inspect the shipping carton for damage. If damage has occurred, contact the carrier and
request a damage report prior to opening the carton. Notation of damages with the delivery
person will speed the filing of damage claims.
CAUTION: Do not lift the MCH by the drain or water connections. Lifting by
!
these connectors will cause equipment damage.
To open the carton, carefully cut the strapping material at the bottom of the carton. Lift and
remove the upper portion of the carton. Lift the unit (per the Lifting instructions) from the base
of the carton and place directly on the floor. Check the carton to be sure that no loose parts or
documents are inside. Store the carton for later shipment of the MCH.

Physical Inspection

1. Check power cord for cuts, exposed wire, bent or missing pins
2. Inspect external cabinet and controls for dents or other physical damage.
3. Check that all warning labels are properly affixed.
4. Inspect external water connectors for leaks or physical damage. Look for water accumulation
on or under the unit, as this may indicate an internal leak.
5. Clean and inspect the water strainers located in the water tank.
6. Have qualified service personnel remove side panels and perform steps 7-9.
WARNING: Electric shock hazard. Access panels should be removed only by
!
qualified service personnel.
7. Inspect the inside of the cabinet for water leaks, loose water connections, and visible damage.
8. Inspect the internal wiring for exposed wires, loose connections, and visible damage.
9. Replace the side panels.

Initial Operation Check

Prior to operation in a clinical setting for the first time, or after extended storage, the following
procedures should be performed. Contact CardioQuip if any failures are encountered.
WARNING: It is the operator's responsibility to use, check, and maintain this
!
device according to the labels of the product, accompanying instruction manuals,
and any revisions of the labeling or instructions that may be subsequently issued.
CardioQuip Modular Cooler-Heater Operator/Service Manual

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents