Troubleshooting

If the unit malfunctions, use the tables below.
Table 11. Patient Preparation Errors
Patient Preparation Error
No signal or low amplitude signal on the Holter
System Preview ECG
Table 12. Holter Recorder Acquisition Errors displayed on the LED (HR 100 only)
Recorder LED will flash yellow several times, pause two seconds, and repeat. Remove
the battery to repeat power-up diagnostics.
Flashes on LED
Table 13. Holter Recorder Acquisition Errors displayed on the LCD (HR 300/1200 only)
LCD
E1
E2
E3
E4
E5
E6
E7
E8
E9
Description
1
Not used
2
Configuration error – recorder SD card contains data or awaiting Bluetooth initialization
3
Low/weak battery
4
Secure digital card not detected
5
Internal data processing error
6
Power on self test failed
7
Wireless data link error
8
Error writing file to secure digital card
9
Error reading file from secure digital card
Description
Not used
Configuration error – recorder SD card contains data or awaiting Bluetooth initialization
Low/weak battery
Secure digital card not detected
Internal data processing error
Power on self test failed
Wireless data link error
Error writing file to secure digital card
Error reading file from secure digital card
Probable Cause and Solution
Date Code on Hook-up kit expired. Check the expiration
date on the Hook-up kit packaging.
Poor skin preparation. Remove sensor(s), repeat skin
prep, select new position (see hookup procedure), and
apply new electrodes.