Welch Allyn HR100 Directions For Use Manual page 15

Holter recorder
Table of Contents

Advertisement

Directions for Use
To Prepare the Patient and Place Electrodes
Help the patient get comfortable. Patient preparation is important for a successful Holter
procedure.
1.
Describe the procedure to the patient.
2. Prepare electrode locations. See
on page 13 for 7-lead placement,
3. Shave the area for electrode placement, if necessary.
4. Clean electrode sites with alcohol.
5. Allow electrode site to dry.
6. Attach the electrodes to the lead wires before attaching to patient.
7 .
Secure each lead wire. Loop each lead wire into a 1- to 2-inch diameter loop, position
it approximately two inches from each sensor site, and tape it to the skin (see
Figure
Figure 3. Patient Lead Wire Stress Loop
Evaluating the Signal Quality from Each Lead
Refer to the Holter System Application Directions for Use.
Clinician tasks
Explain to the patient:
the recording procedure
how to record information in the patient diary
how to use the Patient Event Button
to avoid contact with water
Caution Always ensure that the appropriate electrode placement is employed
for the lead configuration selected.
Caution ECG electrodes can cause skin irritation. Examine the skin for signs of
irritation or inflammation and avoid placement of the electrode in those areas.
3). This reduces movements that cause signal artifact.
Figure 4
on page 12 for 5-lead placement,
Figure 8
on page 15 for 12-lead placement.
11
Figure 5

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hr300Hr1200

Table of Contents