Equipment Overview

General Description

Table 1. CardioLab II Plus Functionality for Different Versions of CardioLab Software
CardioLab II Plus Functionality
Intracardiac
Stimulator
Auxiliary Input
2-2
The CardioLab II Plus Amplifier is designed to amplify various types of
cardiac signals that are acquired from a patient and transmit them back
to the CardioLab/Mac-Lab system, referred to in this manual as the
acquisition computer. These signals include 12 Lead ECG, intra-cardiac
and pressure signals.
The CardioLab II Plus Amplifier is fully supported by CardioLab EP
software versions 3.1 and above and CardioLab Cath software versions
1.0 and above. However, the CardioLab II Plus Amplifier was designed
specifically for use with the CardioLab/Mac-Lab system. The following
table lists the functionality of the CardioLab II Plus Amplifier with the
different software versions.
CardioLab EP 3.1 and Above/CardioLab
Cath 1.0 and Above
Blocks A and B Only
CardioLab II Amplifier Equivalent
Not Available
The CardioLab II Plus Amplifier supports 32, 64, 96 and 128 channel
configurations.
The following list illustrates the number of Catheter Input Modules
(CIM) used with the four configurations of the CardioLab II Plus
Amplifier:
32 Channels: 1 CIM
64 Channels: 3 CIMs
96 Channels: 5 CIMs
128 Channels: 7 CIMs
The Catheter Input Modules for the four configurations of the CardioLab
II Plus Amplifier connect to the following catheter inputs on the front of
the amplifier:
32 Channels: Input A
64 Channels: Inputs A-C
96 Channels: Inputs A-E
128 Channels: Inputs A-G
Cardiolab® II Plus Amplifier Manual
CardioLab/Mac-Lab System
All Blocks Available
Enhanced Switching Options
RF Filtered and Unipolar Inputs
2002938-002E