Recommended separation distances between portable and mobile RF communications
equipment and the device
The device is intended for use in an environment in which radiated RF disturbances are controlled.
The customer or the user of the device can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the device as recommended below, according to the maximum output power of
the communications equipment.
Rated maximum output
power of transmitter
(W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in metres (m) can be determined using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
Notes:
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.

Symbols

The following symbols may appear on the product or packaging.
Read instructions before use.
use.
Manufacturer.
Catalog number.
Protected against finger sized objects and against dripping water when tilted up to 15 degrees
from specified orientation.
Humidity limitation.
control logo 1.
law restricts these devices to sale by or on the order of a physician).
water level.
Use distilled water only.
limitation.
Complies with RTCA DO-160 section 21, category M.
the vicinity of an MRI device).
54
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
d = 0.35 √P
0.035
0.11
0.35
1.1
3.5
Indicates a warning or caution.
European Authorized Representative.
Serial number.
Direct current.
Temperature limitation.
China pollution control logo 2.
80 MHz to 800 MHz
d = 0.35 √P
0.035
0.11
0.35
1.1
3.5
Device number.
Type BF applied part.
Non-ionising radiation.
Prescription only (In the US, Federal
Operating altitude.
800 MHz to 2.5 GHz
d = 0.7 √P
0.070
0.22
0.70
2.2
7.0
Follow instructions before
Batch code.
On / Off.
Device weight.
Class II equipment.
China pollution
Maximum
Atmospheric pressure
MR unsafe (do not use in