Reprocessing

When the device is used for multiple patients, for example, in a sleep lab, clinic, hospital or at a
health care provider, the cleanable water tub, air outlet and air tubing should be reprocessed
between each patient use.
If the cleanable water tub or the air tubing are being used for a single user in the home, refer to the
cleaning instructions in this guide or in the User Guide.
Described here are ResMed's recommended and validated procedures for cleaning and disinfecting
the cleanable water tub, air outlet and air tubing. However, the steps for disinfection vary regionally
and each healthcare facility should consult its own procedures before carrying out those within this
guide.
Note: The standard water tub cannot be disinfected. If contaminated, it must be discarded and
replaced with a new water tub.
WARNING
• ResMed cannot give any assurance that deviations from the procedures listed in this guide,
and their effect on the performance of the product, will be acceptable.
• When using detergents, disinfectants or sterilization agents, always follow the
manufacturer's instructions.
• Beware of electrocution. Do not immerse the device, power supply or power cord in water.
If liquids are spilled into or onto the device, unplug the device and let the parts dry. Always
unplug the device before cleaning and make sure that all parts are dry before plugging it
back in.

Surface disinfection

1. Wipe the exterior of the device including display, externally accessible ports, side cover, power
supply unit and accessories with a disposable cloth and mild detergent or alcohol disinfectant
(see list below).
2. Remove any excess disinfectant with a disposable dry cloth.
Agents recommended for surface disinfection and cleaning:
Warm water and mild detergent eg, Teepol
Window cleaner or other premixed surface detergent
Methyl alcohol solution
70% Ethyl alcohol solution
70-90% Isopropanol solution
10% Bleach solution
Isopropyl wipes
CaviCide
®
Mikrozid
Actichlor
Plus
Terralin
®
.
Note: Agents may not be available in all regions.
32
multipurpose detergent