Comfort - ResMed AirCurve 10 ST-A Clinical Manual

Hide thumbs Also See for AirCurve 10 ST-A:
Table of Contents

Advertisement

Parameter
Description
Trigger
Set the level of
inspiratory flow above
which the device
changes from EPAP to
IPAP.
Cycle
Set the level of
inspiratory flow below
which the device
changes from IPAP to
EPAP.
Height
Sets the body height
needed for the dead
space determination.
Mask
Select the type of mask
used by the patient.
Refer to Mask Device
Compatibility List on
www.resmed.com.

Comfort

Parameter
Description
Ramp Time
Set the ramp time.
Ramp Down
Enable / disable the
Ramp Down feature.
Start
Set the pressure at the
Pressure
start of ramp, up to
treatment pressure.
Start EPAP
Set the pressure at the
start of ramp, up to
minimum treatment
pressure.
Climate Ctrl
Available when HumidAir
humidifier is used and
ClimateLineAir heated air
tubing is connected.
Tube Temp.
Set the minimum
temperature of air
delivered by heated air
tubing such as
ClimateLineAir.
Humidity
Set the humidity level.
Level
20
Mode
S
ST
T
PAC iVAPS CPAP
Mode
S
ST
T
PAC iVAPS CPAP
Range
Very Low / Low / Med / High /
Very High
Very Low / Low / Med / High /
Very High
44–100 inches (110-250 cm),
2-inch increments
Full Face / Nasal / Pillows /
Pediatric
Range
Off / 5–45 mins
Off / On
4–Set pressure, 0.2 cm H
(0.2 hPa) increments
3–EPAP, 0.2 cm H
increments
Manual / Auto
Off / 60–86°F (16–30°C),
1° increments
Off / 1–8
O
2
O (0.2 hPa)
2