Preparing the device for use between patients

When the device is used for multiple patients, for example, in a sleep lab, clinic, hospital or at a health care
provider, the outlet connector, air tubing , cleanable humidifier tub and device enclosure should be
reprocessed between each patient.
Described here are ResMed's recommended and validated procedures for cleaning and disinfecting the
outlet connector, air tubing, cleanable humidifier tub and device enclosure. The person executing
reprocessing activities is responsible for ensuring that reprocessing is completed in line with ResMed's
validated procedures. Components not identified in the reprocessing instructions do not require
reprocessing or are intended for single patient use.
Note: The standard HumidAir 11 tub cannot be reprocessed.
WARNING
Always follow cleaning and disinfection instructions. Some cleaning products may damage the
components and their function or leave harmful residual vapors.
Any deviations from the procedures or claimed maximum number of cycles in this guide can
have an adverse effect on the components and consequently the safety or the quality of therapy.
When using detergents, disinfectants or equipment, always follow the instructions provided by
the manufacturer of those products. In the event of conflict, this guide takes precedence.
Always follow safe operating practices, including the use of appropriate Personal Protective
Equipment (PPE), as required. Refer to the instructions provided by the manufacturer of those
products for more details.
Beware of electrocution:
Do not immerse the device, AC Adaptor or power cord in water.
Do not connect to power while the device is wet. Make sure that all parts are dry before
plugging it in.
General summary
ResMed has validated the following number of cycles for cleaning and disinfection using a manual cleaning
method.
Components
System component
Outlet connector
HumidAir 11 Cleanable tub
ClimateLineAir 11
Standard tubing
SlimLine
Device
Device Enclosure
Cleaning - Mild alkaline, anionic detergent (eg, Alconox)
Chemical high level disinfection
eg, CIDEX-OPA
120
120
30
30
30
Cleaning and low level disinfection
CaviWipes1™
Thermal high level disinfection
167ºF (75ºC) for 30 minutes
80
80
30
30
30